We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We have demonstrated a disposable microfabricated fluorescence-activated cell sorter (microFACS) for sorting various biological entities. Compared with conventional FACS machines, the microFACS provides higher sensitivity, no cross-contamination, and lower cost. We have used microFACS chips to obtain substantial enrichment of micron-sized fluorescent bead populations of differing colors. Furthermore, we have separated Escherichia coli cells expressing green fluorescent protein from a background of nonfluorescent E. coli cells and shown that the bacteria are viable after extraction from the sorting device. These sorters can function as stand-alone devices or as components of an integrated microanalytical chip.

Original publication

DOI

10.1038/15095

Type

Journal article

Journal

Nat Biotechnol

Publication Date

11/1999

Volume

17

Pages

1109 - 1111

Keywords

Escherichia coli, Flow Cytometry, Green Fluorescent Proteins, Luminescent Proteins, Silicones